TRYPTICASE SOY AGAR, 500ML BOTTLE 299099

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-09-17 for TRYPTICASE SOY AGAR, 500ML BOTTLE 299099 manufactured by Becton Dickinson Gmbh.

Event Text Entries

[1430064] This product is a 500ml bottle of agar that is melted then poured into plates or tubes. The technologists in the lab were using a microwave to melt the five bottles of media rather than a water bath. The bottles exploded in the microwave and three techs got hurt. One or more of the following treatments were provided to the individuals involved in this incident: x-rays, stitches, bandage to cover wounds, antibiotics and tylenol. No additional specific treatment information was provided.
Patient Sequence No: 1, Text Type: D, B5


[8751883] This incident is the same event referenced in mdr report #1119779-2010-00005. There were two different products in the microwave when this incident occurred. Both media are labeled "for lab use only". This customer is an industrial testing laboratory that melts the agar in order to form plates/tubes of media for use in specimen testing. The customer also follows usp guidelines. The following is an excerpt from general information (non-mandatory) usp chapter 1117 good laboratory practices regarding the remelting of media: "remelting of an original container of solid media should be performed only once to avoid media whose quality is compromised by overheating or potential contamination. It is recommended that remelting be performed in a heated water bath or by using free-flowing steam. The use of microwave ovens and heating plates is common, but care should be taken to avoid damaging media by overheating and to avoid the potential injury to laboratory personnel from glass breakage and burns. The molten agar medium should be held in a monitored water bath at a temperature of 45 to 50 for not more than 8 hours. " although usp acknowledges the use of a microwave, bd does not recommend this. The following excerpt for handling this type of media appears in the (b)(4) manual, manual of microbiological culture media: "note: use of a microwave oven to melt tubed or bottled media is not recommended. " no corrective actions are planned at this time.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1119779-2010-00004
MDR Report Key1842588
Report Source05
Date Received2010-09-17
Date of Report2010-09-16
Date of Event2010-08-18
Date Mfgr Received2010-08-18
Device Manufacturer Date2010-05-01
Date Added to Maude2010-10-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactCHARLOTTE DANNENFELSER
Manufacturer Street7 LOVETON CIRCLE
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal21152
Manufacturer Phone4103164367
Manufacturer G1BECTON DICKINSON GMBH
Manufacturer StreetTULLASTRASSE 8-12
Manufacturer CityHEIDLEBERG 69126
Manufacturer CountryGM
Manufacturer Postal Code69126
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRYPTICASE SOY AGAR, 500ML BOTTLE
Product CodeJSG
Date Received2010-09-17
Catalog Number299099
Lot Number0117089
Device Expiration Date2011-11-07
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON DICKINSON GMBH
Manufacturer AddressTULLASTRASSE 8-12 HEIDLEBERG 69126 GM 69126


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2010-09-17

© 2026 FDA.report