AMS 700 INFLATABLE PENILE PROSTHESIS 72404324

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-09-03 for AMS 700 INFLATABLE PENILE PROSTHESIS 72404324 manufactured by American Medical Systems, Inc..

Event Text Entries

[20952022] A (b)(6) male was implanted with an ipp device on (b)(6) 2009. Pt measured left and right proximal 12 cm and distal 10cm. The 4. 0cm rtes were used with an 18cm device, measuring 22cm. On (b)(6) 2009, the 4. 0cm rtes were removed and 3. 0cm rtes were implanted. Pt's proximal and distal measurements were re-done measuring both right and left proximal at 11 and distal at 10cm, total measurement equaled 21cm not 22cm. Ams has requested additional info.
Patient Sequence No: 1, Text Type: D, B5


[21202935] It appears the event was not related to a device malfunction but related to the size of the device and corporal measurements. No components were returned for eval. Is sufficiently captured in our risk analysis. Is sufficiently included in the product labeling. Three attempts have been made to obtain additional info and were unsuccessful. Should additional info become available regarding a revision surgery, it will be re-evaluated and a follow-up report sent.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2183959-2010-00336
MDR Report Key1842686
Report Source05
Date Received2010-09-03
Date of Report2009-11-20
Date of Event2009-11-20
Date Mfgr Received2009-11-20
Device Manufacturer Date2009-05-01
Date Added to Maude2010-09-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCAROL SCHMOCK, MDR CONTACT
Manufacturer Street10700 BREN RD., WEST
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal55343
Manufacturer Phone9529306425
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMS 700 INFLATABLE PENILE PROSTHESIS
Generic NameIPP
Product CodeJCW
Date Received2010-09-03
Model Number72404324
Catalog Number72404324
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAMERICAN MEDICAL SYSTEMS, INC.
Manufacturer Address10700 BREN RD., WEST MINNETONKA MN US


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2010-09-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.