SILIKON 1000 OIL 8065601185

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2010-09-02 for SILIKON 1000 OIL 8065601185 manufactured by Alcon - Fort Worth / Alcon Labs Inc..

Event Text Entries

[20745820] Adverse event(s): "no harm" (no consequences or impact to pt). Product problem(s): "milky whitish" (no code available); "too hard to be injected" (physical resistance). A surgeon reported that when the surgeon tried to inject the product, "it was too hard to be injected and exchanged the product to water". When the residue product was exchanged to water, the product turned milky/whitish. The surgeon removed the product from the eye, and instead used a competitor's product. There was no harm to the pt. No further info is expected.
Patient Sequence No: 1, Text Type: D, B5


[20824653] The product was returned for analysis and was verified to have signs of mis-handling. The product has been subjected to a secondary sterilization cycle by the end user. Directions for use in the product insert state that re-sterilization of the product is not to be performed. While we were unable to determine the origin of the damage, our observations reasonably suggest the damage is not mfg related. There have been no other complaints reported in the lot number. No further info is expected. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1610287-2010-00113
MDR Report Key1854135
Report Source01,05
Date Received2010-09-02
Date of Report2010-08-03
Date of Event2010-08-03
Date Mfgr Received2010-08-03
Date Added to Maude2010-11-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHARLES DOLBEE
Manufacturer Street6201 SOUTH FREEWAY
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175518317
Manufacturer G1ALCON - FORT WORTH / ALCON LABS INC
Manufacturer Street6201 SOUTH FREEWAY
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal Code76134
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameSILIKON 1000 OIL
Generic NameFLUID, INTRAOCULAR
Product CodeLWL
Date Received2010-09-02
Returned To Mfg2010-08-24
Model NumberNA
Catalog Number8065601185
Lot NumberTPHU-09-03
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALCON - FORT WORTH / ALCON LABS INC.
Manufacturer Address6201 SOUTH FREEWAY FORT WORTH TX 76134 US 76134


Patients

Patient NumberTreatmentOutcomeDate
10 2010-09-02

© 2026 FDA.report