AFFIRM VPIII AMBIENT TERPERATURE TRANSPORT SY 446255

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-10-01 for AFFIRM VPIII AMBIENT TERPERATURE TRANSPORT SY 446255 manufactured by Becton Dickinson Infusion Therapy Systems.

Event Text Entries

[1477079] Customer was squeezing the atts ampule and a piece of glass pierced the plastic and injured the tech's finger. Basic first aid was administered on (b)(6) 2010 but her finger continued to hurt. Doctor opened the wound and extracted a piece of glass on (b)(6) 2010. She has had no issues since.
Patient Sequence No: 1, Text Type: D, B5


[8771706] This product contains an ampule of reagent. The ampule is crushed and the reagent is released through a dropper into a pt sample. Returns were unavailable for analysis. The batch history record was reviewed and found to be satisfactory. Bd offers a device called "robogrips" which customers use to crush these ampules. A set of robogrips have been sent to the customer with instructions for use. To better educate customers regarding the use of this product, bd has revised the package insert to add a precaution regarding this type of issue and include a reference for use of the robogrip pliers. Bd will continue to monitor this type of issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1119779-2010-00006
MDR Report Key1854678
Report Source05
Date Received2010-10-01
Date of Report2010-09-30
Date of Event2010-08-31
Date Mfgr Received2010-09-09
Date Added to Maude2011-04-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHARLOTTE DANNENFELSER
Manufacturer Street7 LOVETON CIR.
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal21152
Manufacturer Phone4103164367
Manufacturer G1BECTON DICKINSON INFUSION THERAPY SYS
Manufacturer StreetPERIF. LUIS DONALDO COLOSIO #579
Manufacturer CityNOGALES, SONORA 84048
Manufacturer CountryMX
Manufacturer Postal Code84048
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAFFIRM VPIII AMBIENT TERPERATURE TRANSPORT SY
Product CodeLIO
Date Received2010-10-01
Catalog Number446255
Lot Number0116864
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON DICKINSON INFUSION THERAPY SYSTEMS
Manufacturer AddressPERIF. LUIS DONALDO COLOSIO #579 NOGALES, SONORA 84048 MX 84048


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2010-10-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.