ION NITRIDED CAST CUTTER BLADE 0840023000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2010-10-14 for ION NITRIDED CAST CUTTER BLADE 0840023000 manufactured by Stryker Instruments Cork.

Event Text Entries

[1837111] It was reported that while removing a cast using the cast cutter with a ion nitrided cast cutter blade, the patient received a laceration on the thumb and required stitches. Additional information surrounding the event has been requested from the account.
Patient Sequence No: 1, Text Type: D, B5


[8688537] The cast cutter and blade were returned to the manufacturer for evaluation. A cut test found that the blade was loose. The collar had markings indicating that it was seated incorrectly. If the retaining ring is not seated correctly, it is possible that the screw could back up, causing the blade to become loose. Additional information on the details surrounding this event has been requested from the account. This report will be updated when additional information is provided.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616696-2010-00303
MDR Report Key1873031
Report Source06
Date Received2010-10-14
Date of Report2010-09-17
Date of Event2010-08-16
Date Mfgr Received2010-09-17
Date Added to Maude2010-10-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCASEY METZGER
Manufacturer Street4100 EAST MILHAM AVE.
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal49001
Manufacturer Phone2693237700
Manufacturer G1STRYKER INSTRUMENTS CORK
Manufacturer StreetCARRIGTWOHILL BUSINESS & TECHNOLOGY PARK
Manufacturer CityCARRIGTWOHILL, CORK
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameION NITRIDED CAST CUTTER BLADE
Generic NameCAST REMOVAL INSTRUMENT
Product CodeLGG
Date Received2010-10-14
Returned To Mfg2010-09-17
Catalog Number0840023000
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER INSTRUMENTS CORK
Manufacturer AddressCARRIGTWOHILL, CORK EI


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2010-10-14