OLYMPUS VIDEO TELESCOPE A50021A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-09-30 for OLYMPUS VIDEO TELESCOPE A50021A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[1835351] The user facility reported that during an unspecified type of laparoscopy procedure, the user experienced a complete loss of image. The procedure was reportedly completed with a different, but similar device and there was no pt injury. No further detailed info was provided by the user facility.
Patient Sequence No: 1, Text Type: D, B5


[8920075] The device referenced in this report was returned to olympus for eval. The eval confirmed the user's report of image difficulty. The device was tested and there was no image found. The outer tube was bent, had multiple dents and scratches, which likely contributed to the image difficulty. There were dents on the objective lens, and on the video connector. The meniscus lens was cracked. The light guide cable unit was buckled below the protector boot on the control body side, and the contact pins were corroded. The cause of the user's experience was attributed to a damaged outer tube. Based upon the results of the investigation, the cause of the reported phenomenon was likely due to user handling. This report is being filed as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2010-00032
MDR Report Key1880111
Report Source05,06
Date Received2010-09-30
Date of Report2010-09-03
Date of Event2010-09-03
Date Mfgr Received2010-09-03
Date Added to Maude2011-02-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS VIDEO TELESCOPE
Generic NameTELESCOPE
Product CodeFBP
Date Received2010-09-30
Returned To Mfg2010-09-14
Model NumberA50021A
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
10 2010-09-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.