HUDSON NON-REBREATHING MASK W/ SAFETY VENT 1059

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,08 report with the FDA on 2010-10-07 for HUDSON NON-REBREATHING MASK W/ SAFETY VENT 1059 manufactured by Teleflex Medical.

Event Text Entries

[1836872] The event is reported as: the reservoir bag came off after less than 24 hours of use. No patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


[8892302] The device sample was sent to the mfg site for eval. The results of the investigation are from photos and testing of stock product. Eval: method: device history record review. Results: device history record review - lot number unk but lot# 02j1001357 was substituted for review. No issues were found in the review. Pull and heat testing done and compared to photos provided by customer. Some similar condition was observed when compared to the photos. Results: the green tape used to hold the breathing bag is shorter than usual. Conclusions - complaint confirmed. Root causes: heat, flow and green tape length. The mfg team modified the set-up procedure used to set the length of the green tape dispenser. Also, a visual help with the length will be created and this will be used to adjust the green tape dispense green tape length will be created.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004365956-2010-00289
MDR Report Key1881406
Report Source01,06,08
Date Received2010-10-07
Date of Report2010-09-17
Date of Event2010-07-28
Date Mfgr Received2010-09-17
Date Added to Maude2010-11-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL TAGGART, VP
Manufacturer StreetP.O. BOX 12600
Manufacturer CityRTP NC 27709
Manufacturer CountryUS
Manufacturer Postal27709
Manufacturer Phone9194334916
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetAVE. TRANSFORMACION 5954 PARQUE INDUSTRIAL FINSA
Manufacturer CityNUEVO LAREDO 88275
Manufacturer CountryMX
Manufacturer Postal Code88275
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameHUDSON NON-REBREATHING MASK W/ SAFETY VENT
Generic NameNON-REBREATHING MASK
Product CodeCBP
Date Received2010-10-07
Returned To Mfg2010-09-27
Model NumberNA
Catalog Number1059
Lot NumberUNK
ID NumberNA
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressNUEVO LAREDO MX


Patients

Patient NumberTreatmentOutcomeDate
10 2010-10-07

© 2026 FDA.report