TORQUE LIMITING HANDLE/QUICK RELEASE - 6MM HEX COUPLING 321.133

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2010-10-28 for TORQUE LIMITING HANDLE/QUICK RELEASE - 6MM HEX COUPLING 321.133 manufactured by Synthes Monument.

Event Text Entries

[19686067] During an arcofix procedure, the surgeon was final tightening the gold locking screws with the torque limiting handle/quick release and the head of the screw sheared off in the center position. The head was retrieved with the shaft remaining in the bone and plate. Pt was augmented with posterior pedicle screw fixation and bone graft. The procedure was completed, surgeon noted the plate was locked in place. Torque limiting handle/quick release found to be out of specification due to lack of calibration.
Patient Sequence No: 1, Text Type: D, B5


[19747483] This info was not provided during initial report. Additional info has been requested. Subject device is an instrument and is not implanted. Investigation is on going; no conclusion can be drawn as no device was returned. Review of the manufacturing records has been requested.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1719045-2010-00317
MDR Report Key1888451
Report Source05,07
Date Received2010-10-28
Date of Report2010-09-28
Date of Event2010-07-05
Date Mfgr Received2010-09-28
Device Manufacturer Date2009-07-07
Date Added to Maude2011-02-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHARISSE BALL
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone8006207025
Manufacturer G1SYNTHES MONUMENT
Manufacturer Street1101 SYNTHES AVENUE
Manufacturer CityMONUMENT CO 80132
Manufacturer CountryUS
Manufacturer Postal Code80132
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTORQUE LIMITING HANDLE/QUICK RELEASE - 6MM HEX COUPLING
Generic NameTORQUE LIMITING HANDLE
Product CodeGFC
Date Received2010-10-28
Model NumberNA
Catalog Number321.133
Lot Number596692G08
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSYNTHES MONUMENT
Manufacturer AddressMONUMENT CO 80132 US 80132


Patients

Patient NumberTreatmentOutcomeDate
10 2010-10-28

© 2026 FDA.report