QUINTON TM55 TREADMILL TM55LXXRCE NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-10-21 for QUINTON TM55 TREADMILL TM55LXXRCE NA manufactured by Cardiac Science Corporation.

Event Text Entries

[1664194] During stage one of a stress test protocol, customer stated the treadmill sped up on its own and the patient was thrown from the treadmill. The patient received a minor scrape on the knee.
Patient Sequence No: 1, Text Type: D, B5


[8920292] Field service engineer was dispatched to the site and checked the entire system including internal connections within the treadmill and stress monitor. System log files were also gathered and sent in for evaluation. Several errors related to the treadmill were observed. As a precaution the field service engineer replaced the main treadmill circuit board and drive motor. The treadmill circuit board was returned to cardiac science corporation for investigation. It was installed in a test fixture and tested extensively with no problems found. All tests passed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3014398-2010-00033
MDR Report Key1889177
Report Source05
Date Received2010-10-21
Date of Report2010-09-28
Date of Event2010-09-28
Date Mfgr Received2010-09-28
Device Manufacturer Date2009-09-01
Date Added to Maude2011-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBEV MAGRANE
Manufacturer Street3303 MONTE VILLA PARKWAY
Manufacturer CityBOTHELL WA 98028
Manufacturer CountryUS
Manufacturer Postal98028
Manufacturer Phone4254022365
Manufacturer G1CARDIAC SCIENCE CORPORATION
Manufacturer Street500 BURDICK PARKWAY
Manufacturer CityDEERFIELD WI 53531
Manufacturer CountryUS
Manufacturer Postal Code53531
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameQUINTON TM55 TREADMILL
Generic NameCARDIAC STRESS TESTING TREADMILL
Product CodeIOL
Date Received2010-10-21
Returned To Mfg2010-10-14
Model NumberTM55LXXRCE
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARDIAC SCIENCE CORPORATION
Manufacturer Address3303 MONTE VILLA PARKWAY BOTHELL WA 98028 US 98028


Patients

Patient NumberTreatmentOutcomeDate
10 2010-10-21