3600E HBO CHAMBER, 36IN, ELECTRONIC CONTROL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2010-10-26 for 3600E HBO CHAMBER, 36IN, ELECTRONIC CONTROL manufactured by Sechrist Industries, Inc..

Event Text Entries

[1664705] Patient's first treatment and was taught how to clear ears. Twelve minutes into treatment, going down to pressure and compressing at 1. 0psi/minute heading toward 2ata. When the chamber got to 1. 52ata, it suddenly went back up to 1. 20ata. Then it went from 1. 20ata back to 1. 52ata in 5 seconds. This appeared to cause ear pain to the pt as the pt grabbed her head. Attendant set the pressure dow to 1. 10ata to stop the chamber pressure from elevating. The new set pressure was at 1. 10ata on the dial and the pressure continued going up. It finally stopped at 1. 57ata and then started coming down at a normal rate, slowly, at 1psi/minute. The doctor checked the pt after the pt was removed from the chamber and said pt had a teeds reading of t3 on the right ear.
Patient Sequence No: 1, Text Type: D, B5


[8914086] Device failure related to loss of pressure at user facility. Device requires 45 to 70 psi pressure. Pressure reading at the main gauge read 40 psi.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2020676-2010-00030
MDR Report Key1889819
Report Source06
Date Received2010-10-26
Date of Report2010-10-26
Date of Event2010-09-27
Date Facility Aware2010-09-27
Date Mfgr Received2010-09-27
Date Added to Maude2011-02-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street4225 EAST LA PALMA AVE.
Manufacturer CityANAHEIM CA 92807
Manufacturer CountryUS
Manufacturer Postal92807
Manufacturer Phone7145798400
Manufacturer G1SECHRIST INDUSTRIES, INC.
Manufacturer Street4225 EAST LA PALMA AVE.
Manufacturer CityANAHEIM CA 92807
Manufacturer CountryUS
Manufacturer Postal Code92807
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3600E HBO CHAMBER, 36IN, ELECTRONIC CONTROL
Generic Name3600E
Product CodeCBF
Date Received2010-10-26
Model Number3600E
Catalog Number3600E
OperatorHEALTH PROFESSIONAL
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerSECHRIST INDUSTRIES, INC.
Manufacturer AddressANAHEIM CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2010-10-26