I-STAT CHEM 8+N CARTRIDGE 03P91-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-11-03 for I-STAT CHEM 8+N CARTRIDGE 03P91-25 manufactured by Abbott Point Of Care Canada Ltd.

Event Text Entries

[20414750] On (b)(6) 2010, abbott point of care was contacted by a customer who reported the i-stat chem 8+ cartridge yielded a hematocrit result of 37% pcv. A separate sample, drawn at the same time as the i-stat sample, was tested on the lab instrument and the hematocrit result was 15. 4% pcv. No repeat labs performed. Patient was not treated on either hematocrit result. Patient died of respiratory failure at 6:30pm (b)(6) 2010.
Patient Sequence No: 1, Text Type: D, B5


[20594953] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2245578-2010-00141
MDR Report Key1893966
Report Source05
Date Received2010-11-03
Date of Report2010-11-03
Date of Event2010-10-11
Date Mfgr Received2010-10-14
Device Manufacturer Date2010-06-01
Date Added to Maude2011-03-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactPETER SCOTT
Manufacturer Street400 COLLEGE RD
Manufacturer CityPRINCETON NJ 08540
Manufacturer CountryUS
Manufacturer Postal08540
Manufacturer Phone6094549270
Manufacturer G1ABBOTT POINT OF CARE CANADA LTD
Manufacturer Street185 CORKSTOWN RD
Manufacturer CityOTTAWA, ONTARIO K2H8V4
Manufacturer CountryCA
Manufacturer Postal CodeK2H 8V4
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameI-STAT CHEM 8+N CARTRIDGE
Generic NameCHEM 8+ CARTRIDGE
Product CodeGKF
Date Received2010-11-03
Catalog Number03P91-25
Lot NumberG10170
Device Expiration Date2010-12-14
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT POINT OF CARE CANADA LTD
Manufacturer AddressOTTAWA, ONTARIO CA


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.