COBAS 6000 C501 MODULE 05036453001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-11-17 for COBAS 6000 C501 MODULE 05036453001 manufactured by Roche Diagnostics.

Event Text Entries

[1678187] The user received questionable nh3l (ammonia) results for one patient sample. The original sample was poured into a false bottom tube due to a low sample volume and generated a result of 151 ug/ml. This result was reported outside the laboratory. The same sample in the same false bottom tube was then tested on cobas c501 serial number (b)(4) with a result of 18 ug/ml. The same sample in the same false bottom tube was then put back on the original analyzer and generated a result of 27 ug/ml. This result was sent as a corrective report. The patient was not adversely affected. The ammonia reagent lot number was 61981001. The field service representative determined the cause was unknown. He performed precision and accuracy checks and stated the unit was operational with no errors.
Patient Sequence No: 1, Text Type: D, B5


[8975066] .
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2010-06785
MDR Report Key1901039
Report Source05,06
Date Received2010-11-17
Date of Report2010-11-17
Date of Event2010-11-02
Date Mfgr Received2010-11-02
Date Added to Maude2010-11-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA ERIC KOLODZIEJ
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175212834
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS 6000 C501 MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJIF
Date Received2010-11-17
Model NumberNA
Catalog Number05036453001
Lot NumberNA
ID NumberNA
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.