DIMENSION? CLINICAL CHEMISTRY SYSTEM K5088

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-11-17 for DIMENSION? CLINICAL CHEMISTRY SYSTEM K5088 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[1651010] A falsely elevated acetaminophen result was obtained on a patient sample. The result was reported to the physician. The sample was repeated and a negative result was obtained. The patient was admitted and treated. There was no report of adverse health consequences as a result of the falsely elevated acetaminophen result.
Patient Sequence No: 1, Text Type: D, B5


[8901550] Analysis of the instrument and instrument data indicate that the cause for the falsely elevated acetaminophen result was user error for using an incorrect diluent (water) for diluting the specimen. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2517506-2010-00128
MDR Report Key1901725
Report Source05,06
Date Received2010-11-17
Date of Report2010-11-03
Date of Event2010-11-03
Date Mfgr Received2010-11-03
Device Manufacturer Date2009-10-20
Date Added to Maude2011-04-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. CAROLYN CHASTAIN
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318789
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE
Manufacturer CityNEWARK DE 19702
Manufacturer CountryUS
Manufacturer Postal Code19702
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION? CLINICAL CHEMISTRY SYSTEM
Generic NameTROPONIN I FLEX? REAGENT CARTRIDGE
Product CodeLDP
Date Received2010-11-17
Catalog NumberK5088
Lot Number10181AE
Device Expiration Date2011-06-30
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE NEWARK DE 19702 US 19702


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.