DETECTO SCALE 6745 62450-300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2010-11-09 for DETECTO SCALE 6745 62450-300 manufactured by Cardinal Health.

Event Text Entries

[1672982] Baby scale readings inaccurate by 5-10% wt differences. Biomed has calibrated and attempted to remedy the issue, and when the scale reads their test using a 10 pound weight it reads accurately, but when weighing a moving infant the weights are consistently off by 5-10% of the actual weight. ====================== health professional's impression======================scale does not read accurately and consistently especially with infant moving intermittently.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1901739
MDR Report Key1901739
Date Received2010-11-09
Date of Report2010-11-09
Date of Event2010-11-07
Report Date2010-11-09
Date Reported to FDA2010-11-09
Date Added to Maude2010-11-17
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameDETECTO SCALE
Generic NameSCALE, INFANT
Product CodeFRW
Date Received2010-11-09
Model Number6745
Catalog Number62450-300
Lot Number*
ID Number*
OperatorNURSE
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerCARDINAL HEALTH
Manufacturer Address203 EAST DAUGHERTY WEBB CITY MO 64870 US 64870


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.