HOUSE STAPES CURETTE, STRONG AU12452

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-10-28 for HOUSE STAPES CURETTE, STRONG AU12452 manufactured by Carefusion.

Event Text Entries

[1651505] Broken tip customer states the cup end of the instrument fell in the wound. The piece was retrieved from patient.
Patient Sequence No: 1, Text Type: D, B5


[8973708] The complete effected instrument was not received for evaluation. The complete instrument is crucial for a complete and accurate analysis of the concern. A metallurgical review of the portion received shows that the instrument had been structurally damaged previous to this use and should have been taken out of service. It is not possible to say what caused the previous damage as it may have been due to one of the following: normal wear, misused or even being dropped during a simple cleaning process. The instrument segment provided meets all applicable specifications that would indicate that the failure is not due to workmanship or materials. A review of trends over 10 years and device history has revealed no previous incidents of any nature. The result of this investigation is complete at this time. Due to the findings of this investigation, a corrective and preventive action will not be initiated.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1423507-2010-00043
MDR Report Key1902963
Report Source05
Date Received2010-10-28
Date of Report2010-10-27
Date of Event2010-09-09
Date Mfgr Received2010-09-09
Date Added to Maude2011-04-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMRS. MICHELE DONATICH
Manufacturer Street1430 WAUKEGAN ROAD
Manufacturer CityMCGAW PARK IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8478876412
Manufacturer G1CAREFUSION 2200, INC
Manufacturer Street1500 WAUKEGAN ROAD
Manufacturer CityWAUKEGAN IL 60085
Manufacturer CountryUS
Manufacturer Postal Code60085
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHOUSE STAPES CURETTE, STRONG
Generic NameNA / JYG
Product CodeJYG
Date Received2010-10-28
Returned To Mfg2010-10-05
Model NumberAU12452
Catalog NumberAU12452
Lot NumberA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer AddressMCGAW PARK IL US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2010-10-28

© 2024 FDA.report