UNICEL? DXC 600 PRO SYNCHRON? CHEMISTRY ANALYZER A11810

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-11-20 for UNICEL? DXC 600 PRO SYNCHRON? CHEMISTRY ANALYZER A11810 manufactured by Beckman Coulter, Inc..

Event Text Entries

[1802951] A customer contacted beckman coulter inc (bci) in regards to erroneously low total protein results generated by unicel dxc 600 pro synchron chemistry analyzer. The erroneous results were reported out of the laboratory. The customer noticed the low results and repeated all patient samples since the last good qc on an alternate instrument. Seven patient results were amended. The customer does not believe that patients were impacted by the erroneous results.
Patient Sequence No: 1, Text Type: D, B5


[8975671] Qc did not demonstrate an issue. A bci field service engineer (fse) identified sample probe carryover, and replaced sample probe and wash collar. The fse performed alignments, and a retest passed specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2010-01236
MDR Report Key1903791
Report Source05
Date Received2010-11-20
Date of Report2010-11-19
Date of Event2010-10-22
Date Mfgr Received2010-10-22
Device Manufacturer Date2006-05-16
Date Added to Maude2012-01-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Remedial ActionIN
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameUNICEL? DXC 600 PRO SYNCHRON? CHEMISTRY ANALYZER
Generic NameCLINICAL CHEMSTRY ANALYZER
Product CodeCEK
Date Received2010-11-20
Model NumberDXC 600 PRO
Catalog NumberA11810
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 SOUTH KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.