UNICEL? DXC 800 PRO SYNCHRON? CHEMISTRY ANALYZER A11812

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-11-27 for UNICEL? DXC 800 PRO SYNCHRON? CHEMISTRY ANALYZER A11812 manufactured by Beckman Coulter, Inc..

Event Text Entries

[1801173] A customer contacted beckman coulter inc. (bci) in regards to an erroneously low total bilirubin (tbil) result generated by unicel dxc 800 pro synchron chemistry analyzer. The result was not reported out of the laboratory. Subsequent testing on an alternate analyzer produced higher result. There was no effect to the patient or user.
Patient Sequence No: 1, Text Type: D, B5


[8904508] The sample was analyzed from the primary tube. Qc was trending downwards. No other system errors were noted and the customer confirmed no other chemistries were affected. On (b)(4) 2010, a bci field service engineer (fse) replaced the reagent and sample syringes. The fse found 3 broken cuvettes, glass in one wash/vacuum probe, and vacuum leak at wash manifold. The fse replaced the whole wash manifold assembly, and performed alignments. The fse calibrated both tbil and dbil (direct bilirubin) and performed 20 rep precision tests on both chemistries. All results were within the specifications. The root cause for this event was hardware. This report is linked to medwatch reports #2050012-2010-01421 and 2050012-2010-01422.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2010-01419
MDR Report Key1908440
Report Source05
Date Received2010-11-27
Date of Report2010-10-29
Date of Event2010-10-28
Date Mfgr Received2010-10-29
Device Manufacturer Date2007-12-07
Date Added to Maude2012-01-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Remedial ActionRL
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameUNICEL? DXC 800 PRO SYNCHRON? CHEMISTRY ANALYZER
Generic NameCLINICAL CHEMSTRY ANALYZER
Product CodeJFM
Date Received2010-11-27
Model NumberDXC 800 PRO
Catalog NumberA11812
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 SOUTH KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.