[1671350] During a prodisc-l procedure at l5-s1, the surgeon was using the chisel 10mm and it split the vertebral body and pushed a small piece of the posterior cortex back in towards the spinal cord. Surgeon placed the prodisc-l and completed the procedure. The following day, the patient was returned to the operating room for removal of the small piece of bone. Surgeon indicated he felt the event was due to a dull chisel.
Patient Sequence No: 1, Text Type: D, B5


[8803611] Additional information has been requested. Subject device is an instrument and is not implanted. Investigation could not be concluded, no conclusion could be drawn. No device was returned and no lot number was provided. Manufacturing records could not be reviewed without a lot number. Manufacture date could not be determined without a lot number.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1719045-2010-00339
MDR Report Key1909906
Report Source05,07
Date Received2010-11-17
Date of Report2010-10-29
Date of Event2010-10-28
Date Mfgr Received2010-10-29
Date Added to Maude2010-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJODI TEMPLE
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone8006207025
Manufacturer G1SYNTHES MONUMENT
Manufacturer Street1101 SYNTHES AVENUE
Manufacturer CityMONUMENT CO 80132
Manufacturer CountryUS
Manufacturer Postal Code80132
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHISEL 10MM
Generic NameCHISEL
Product CodeEML
Date Received2010-11-17
Model NumberNA
Catalog NumberPDL322
Lot NumberNI
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSYNTHES MONUMENT
Manufacturer AddressMONUMENT CO 80132 US 80132


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2010-11-17