OLYMPUS VIDEO TELESCOPES A50021A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-11-18 for OLYMPUS VIDEO TELESCOPES A50021A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[1650827] The user facility reported that during a therapeutic lap-appendectomy procedure, the users experienced a complete loss of image. The procedure was completed using a different, but similar device. There was no pt harm reported.
Patient Sequence No: 1, Text Type: D, B5


[8800999] Olympus followed up on this report and was informed that the users had reportedly experienced intermittent image issues in the days prior to the event, and had elected to use the device. The user facility returned the device to olympus for eval. The eval confirmed the user's claim of a complete image loss. The eval found that the outer tube was bent and dented, and the contact pins on the video connector were damaged and appeared burnt. Corrosion was present around the switch unit. In addition, the light guide tube could not be removed and there were loose light guide posts, and dents on the light guide cable. The device was serviced and returned to the user facility. This report is being submitted as an mdr in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2010-00042
MDR Report Key1911493
Report Source05,06
Date Received2010-11-18
Date of Report2010-11-02
Date of Event2010-11-01
Date Mfgr Received2010-10-26
Date Added to Maude2012-01-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS VIDEO TELESCOPES
Generic NameVIDEO TELESCOPE
Product CodeFBP
Date Received2010-11-18
Returned To Mfg2010-11-02
Model NumberA50021A
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-18

© 2026 FDA.report