[1804605] An employee was attempting to make this media in a tightly capped erlenmeyer flask. While heating, the employee was distracted and forgot to loosen the cap. The flask shattered and glass shards cut the employee. The wound required stitches. The contact stated that in the future, the cap will be vented or loosened prior to applying heat.
Patient Sequence No: 1, Text Type: D, B5


[8797833] Desoxycholate agar requires rehydration from dehydrated culture media. Directions to prepare the agar are to blend the required amount of powder and water. Mixing thoroughly. Heat with frequent agitation and boil for 1 minute to completely dissolve the powder. Avoid overheating and do not autoclave. It is standard laboratory practice for the preparation of media to ensure that flasks being heated are properly vented. If heating continues and the pressure is not adequately relieved, the pressure and temperature can rise to a point that the flask may burst. In this instance, the customer admits to improperly venting the container while heating. No further action will be taken.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1119779-2010-00007
MDR Report Key1911508
Report Source05
Date Received2010-11-24
Date of Report2010-11-23
Date of Event2010-10-28
Date Mfgr Received2010-10-28
Date Added to Maude2011-04-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHARLOTTE DANNENFELSER
Manufacturer Street7 LOVETON CIR.
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal21152
Manufacturer Phone4103164367
Manufacturer G1BECTON DICKINSON
Manufacturer Street39 LOVETON CIR.
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal Code21152
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIFCO DESOXYCHOLATE AGAR
Product CodeJSG
Date Received2010-11-24
Catalog Number227310
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBD DIAGNOSTIC SYSTEMS
Manufacturer Address39 LOVETON CIRCLE SPARKS MD 21152 US 21152


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2010-11-24