4 HOLE PIN CLAMP, BLUE MONOTUBE TRIAX 20MM 5150-3-070

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2010-11-17 for 4 HOLE PIN CLAMP, BLUE MONOTUBE TRIAX 20MM 5150-3-070 manufactured by Stryker Osteosynthesis Selzach.

Event Text Entries

[1654012] On (b)(6) 2010, the patient underwent the surgery with the monotube triax (blue) for femoral distraction osteogenesis. The surgeon was monitoring for the patient. The surgeon confirmed that the screw for the fixation of pin clamp and tube were loose. The screw broke when the surgeon tightened the loose screw with the long handle wrench. The surgeon changed the monotube triax to another monotube triax on (b)(6) 2010. The broken monotube triax (blue) bought and was the second use. The surgeon requested the investigation of the event and the improvement of wrench.
Patient Sequence No: 1, Text Type: D, B5


[8824158] Additional information has been requested and if received will be provided on a supplemental report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8031020-2010-00157
MDR Report Key1915334
Report Source01,07
Date Received2010-11-17
Date of Report2010-10-29
Date of Event2010-10-28
Date Mfgr Received2010-10-29
Date Added to Maude2012-01-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactROSE MINCIELI
Manufacturer Street325 CORPORATE DR.
Manufacturer CityMAHWAH NJ 07430
Manufacturer CountryUS
Manufacturer Postal07430
Manufacturer Phone2018315000
Manufacturer G1STRYKER OSTEOSYNTHESIS SELZAH
Manufacturer StreetBOHNACKERWEG 1
Manufacturer CitySELZACH 2545
Manufacturer CountrySZ
Manufacturer Postal Code2545
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand Name4 HOLE PIN CLAMP, BLUE MONOTUBE TRIAX 20MM
Generic NameIMPLANT
Product CodeKIT
Date Received2010-11-17
Model NumberNA
Catalog Number5150-3-070
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER OSTEOSYNTHESIS SELZACH
Manufacturer AddressSELZACH 2545 SZ 2545


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.