GAMBRO

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2010-11-29 for GAMBRO manufactured by Gambro Americas Headquarters.

Event Text Entries

[16920288] Machine alarms during the disinfection process. Alarm occurs during the rinse following the bleach, consistently when the clock counts down to 12:58. Ld1 level sensor error # 33 and time out process # 190.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5018410
MDR Report Key1916847
Date Received2010-11-29
Date of Report2010-11-29
Date of Event2010-06-04
Date Added to Maude2010-12-10
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameGAMBRO
Generic NamePHOENIX DIALYSIS MACHINE
Product CodeFII
Date Received2010-11-29
Lot NumberPE 35804848
ID NumberPH 13668
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerGAMBRO AMERICAS HEADQUARTERS
Manufacturer Address14143 DENVER W PKWY LAKEWOOD CO 80401 US 80401


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-29

© 2026 FDA.report