GAMBRO

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2010-11-29 for GAMBRO manufactured by Gambro Americas Headquarters.

Event Text Entries

[17722948] Arterial gauge side reads 0; it is not registering when pt is on machine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5018411
MDR Report Key1916868
Date Received2010-11-29
Date of Report2010-11-29
Date of Event2010-08-11
Date Added to Maude2010-12-10
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameGAMBRO
Generic NamePHOENIX DIALYSIS MACHINE
Product CodeFII
Date Received2010-11-29
Lot NumberPE 35804848
ID NumberPH #13668
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerGAMBRO AMERICAS HEADQUARTERS
Manufacturer Address14143 DENVER WEST PKWY LAKEWOOD CO 80401 US 80401


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.