COBAS 6000 E601 MODULE 04745922001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2010-12-15 for COBAS 6000 E601 MODULE 04745922001 manufactured by Roche Diagnostics.

Event Text Entries

[8945022] A specific root cause was not identified. A general instrument problem, assay problem, or handling problem could not be identified. Results were not reported outside the laboratory or to the physician.
Patient Sequence No: 1, Text Type: N, H10


[19932189] The customer received questionable cortisol results for one patient sample. The same sample was tested six times during three different days. The initial results were 49 nmol/l, 77 nmol/l and 91 nmol/l. The sample was repeated on (b)(6) 2010 and recovered 94 nmol/l and 119 nmol/l. The sample was repeated on (b)(6) 2010 and recovered 113 nmol/l. Results were not reported outside the laboratory and were not reported to the physician. The laboratory physician contacted the patient's doctor. The patient was not harmed by any action taken and the patient's current condition is good. The cortisol reagent lot number was 15879801.
Patient Sequence No: 1, Text Type: D, B5


[20098575] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2010-07454
MDR Report Key1927172
Report Source01,05,06
Date Received2010-12-15
Date of Report2011-01-04
Date of Event2010-11-28
Date Mfgr Received2010-12-03
Date Added to Maude2011-04-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA ERIC KOLODZIEJ
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175212834
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS 6000 E601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeCGR
Date Received2010-12-15
Model NumberNA
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2010-12-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.