[1686756] The gti test kit is used to determine hit antibodies. However, it has been learned that the test kit is giving false positives as confirmed by the blood ctr of (b)(6). Gti does not feel this is clinically significant and recommends if there is a question to purchase gti's confirmation kits. However, the question then becomes what do you do if you get what you believe is a false positive, use gti's confirmation kit and get a negative result. This is not a satisfactory solution where it is believed there is a product giving false positives. Our blood bank physician can provide much more detailed info if desired. The local solution is that if gti's test is negative, then no further action. If there is a positive test using gti's kit, another sample will be sent to the blood ctr of (b)(6) for verification. This is not an optimal solution but the only one we feel comfortable with at the present time. Diagnosis or reason for use: hit testing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5018711
MDR Report Key1942098
Date Received2010-12-23
Date of Report2010-12-23
Date of Event2010-12-17
Date Added to Maude2011-01-12
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameGTI TEST KITS
Generic NameGTI TEST KITS
Product CodeMYP
Date Received2010-12-23
Lot Number031710-X45
Device Expiration Date2012-02-03
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerGTI DIAGNOSTICS


Patients

Patient NumberTreatmentOutcomeDate
10 2010-12-23