PRISM NUCLEAR GAMMA CAMERA SYSTEM 453566512911 88201

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-11-16 for PRISM NUCLEAR GAMMA CAMERA SYSTEM 453566512911 88201 manufactured by Philips Medical Systems (cleveland), Inc..

Event Text Entries

[18490105] A customer reported a monitor failure at the gantry control which was followed by smoke and a burning smell from the capacitor in the power supply. The defective power supply has been determined to be responsible for the smoke development. A new power supply has been ordered for replacement. The fire dept was called.
Patient Sequence No: 1, Text Type: D, B5


[18698728] Note: we have not completed our investigation of this event. We will file a follow-up mdr at the completion of the investigation. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1525965-2010-00037
MDR Report Key1942474
Report Source05,06
Date Received2010-11-16
Date of Report2010-10-18
Date Mfgr Received2010-10-18
Date Added to Maude2011-08-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactELIZABETH MCLAIN
Manufacturer Street5520 NOBEL DR.
Manufacturer CityFITCHBURG WI 53711
Manufacturer CountryUS
Manufacturer Postal53711
Manufacturer Phone6082886948
Manufacturer Street595 MINER RD.
Manufacturer CityCLEVELAND OH 44143
Manufacturer CountryUS
Manufacturer Postal Code44143
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRISM NUCLEAR GAMMA CAMERA SYSTEM
Product CodeIYX
Date Received2010-11-16
Model Number453566512911
Catalog Number88201
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
Manufacturer Address595 MINER RD. CLEVELAND OH 44143 US 44143


Patients

Patient NumberTreatmentOutcomeDate
10 2010-11-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.