ELITE AND USA COLD KNIFE, STRAIGHT BLADE, 3/PKG K-SB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2010-12-28 for ELITE AND USA COLD KNIFE, STRAIGHT BLADE, 3/PKG K-SB manufactured by Gyrus Acmi Inc..

Event Text Entries

[14883054] During a procedure, while using the elite cold knife straight blade, the blade broke off during use in the patient's bladder. The blade was successfully retrieved without any injury to the patient.
Patient Sequence No: 1, Text Type: D, B5


[15167240] Visual examination of the returned portion (blade only) of the k-sb instrument finds that the blade has broken cleanly from the base of the instrument. The blade exhibits a break profile on the right side of the blade that suggests side loading during use caused the blade to fracture. The fracture originated on the side of the blade, indicating side loading. This device is intended for axial cutting only. The instruction for use manual warns against side loading in the warnings section. The device failed due to abnormal use by the customer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006159227-2010-00006
MDR Report Key1948682
Report Source06
Date Received2010-12-28
Date of Report2010-12-28
Date of Event2010-11-29
Date Facility Aware2010-11-29
Date Mfgr Received2010-11-29
Date Added to Maude2012-03-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactTERRENCE SULLIVAN
Manufacturer Street136 T7URNPIKE ROAD
Manufacturer CitySOUTHBOROUGH MA 01772
Manufacturer CountryUS
Manufacturer Postal01772
Manufacturer Phone5088042739
Manufacturer G1GYRUS ACMI INC.
Manufacturer StreetCARRETERA SALTILLO-ZACATECAS KM 4.5 LA ANGOSTURA C.P.
Manufacturer CitySALTILLO, COAHUILA 25000
Manufacturer CountryMX
Manufacturer Postal Code25000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELITE AND USA COLD KNIFE, STRAIGHT BLADE, 3/PKG
Generic NameCOLD KNIFE, STRAIGHT BLADE, 3/PKG
Product CodeEZO
Date Received2010-12-28
Returned To Mfg2010-12-22
Model NumberK-SB
Catalog NumberK-SB
Lot NumberH883KSB07
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeNA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI INC.
Manufacturer AddressCARRETERA SALTILLO-ZACATECAS KM 4.5 LA ANGOSTURA., CP MX


Patients

Patient NumberTreatmentOutcomeDate
10 2010-12-28

© 2026 FDA.report