MICROALBUMIN OSR6167

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2011-01-05 for MICROALBUMIN OSR6167 manufactured by Beckman Coulter Biomedical Limited.

Event Text Entries

[1719771] A customer contacted beckman coulter inc. (bci) in regards to erroneous microalbumin results generated on au5400 chemistry analyzer for two patients. The results were not reported out of the laboratory. Subsequent testing produced lower results. There was no effect to the patients or user.
Patient Sequence No: 1, Text Type: D, B5


[8814765] The presence of drugs may have interfered with the microalbumin tests, and produced excessively high results with flags. The system was configured in such a way that the flags triggered an automatic rerun with a 1/10 dilution. This dilution may have reduced the background absorbance to tolerable levels, but also diluted the urine albumin concentration below the bottom of the dynamic range, giving results with g flags which were then multiplied by the dilution ratio (10) to give a final result apparently inside the measuring range (5 - 300 mg/l). The samples are not available for further investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-00001
MDR Report Key1949859
Report Source05
Date Received2011-01-05
Date of Report2010-12-07
Date of Event2010-12-07
Date Mfgr Received2010-12-07
Date Added to Maude2011-05-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER BIOMEDICAL LIMITED
Manufacturer StreetLISMEEHAN O'CALLAGHAN'S MILLS
Manufacturer CityCO. CLARE,
Manufacturer CountryEI
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameMICROALBUMIN
Generic NameMICROALBUMIN TEST
Product CodeJIR
Date Received2011-01-05
Model NumberNA
Catalog NumberOSR6167
OperatorHEALTH PROFESSIONAL
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER BIOMEDICAL LIMITED
Manufacturer AddressLISMEEHAN O'CALLAGHAN'S MILLS CO. CLARE, EI


Patients

Patient NumberTreatmentOutcomeDate
10 2011-01-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.