XIMATRON C RADIOTHERAPY TREATMENT SIMULATOR XIM-CX

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2011-01-06 for XIMATRON C RADIOTHERAPY TREATMENT SIMULATOR XIM-CX manufactured by Varian Medical Systems Uk Ltd.

Event Text Entries

[15207094] Treatment fields were deleted by the system during simulation. Upon saving the images, a system warning message appeared, stating that one according field would not exist for one of the image geometry and was deleted. No other errors or messages.
Patient Sequence No: 1, Text Type: D, B5


[15794116] No serious injury was reported as a result of this malfunction. It is believed unlikely, however, if the field were deleted and the operator did not notice, and no subsequent practices caught the error, treatment could result in an underdose. Though still under investigation, varian has determined that a mdr is appropriate as this malfunction, should it recur, could potentially cause serious injury. Additional follow-up to this mdr is expected upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8020711-2010-00010
MDR Report Key1961444
Report Source01,05
Date Received2011-01-06
Date of Report2010-12-21
Date of Event2010-11-15
Date Mfgr Received2010-12-21
Device Manufacturer Date1997-02-01
Date Added to Maude2012-01-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL PIGNATARO, MANAGER
Manufacturer Street911 HANSEN WAY, M/S C-255
Manufacturer CityPALO ALTO CA 84304
Manufacturer CountryUS
Manufacturer Postal84304
Manufacturer Phone6504246471
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXIMATRON C RADIOTHERAPY TREATMENT SIMULATOR
Generic NameSYSTEM, SIMULATION, RADIATION THERAPY
Product CodeKPQ
Date Received2011-01-06
Model NumberXIM-CX
ID Number8.2.24
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVARIAN MEDICAL SYSTEMS UK LTD


Patients

Patient NumberTreatmentOutcomeDate
10 2011-01-06

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.