THEME SERIES INFANT SCALE/TABLE 7830

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2011-02-03 for THEME SERIES INFANT SCALE/TABLE 7830 manufactured by Clinton Industries Inc.

Event Text Entries

[18613524] The screws that attached the fall protection plexi-glass end on an infant scale are stripping out. A sibling of a pt in the pediatrics clinic was seen with the screws in his hand. Potential choking and fall hazard. Medical maintenance inspected all the infant scales in our facility out of the 11 scales medical maintenance identified and repaired on the spot 10 scales. Dates of use: (b)(6) 2008 to present.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5019339
MDR Report Key1985386
Date Received2011-02-03
Date of Report2010-12-23
Date of Event2010-12-21
Date Added to Maude2011-02-15
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameTHEME SERIES INFANT SCALE/TABLE
Generic NameINFANT SCALE
Product CodeFRW
Date Received2011-02-03
Model Number7830
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerCLINTON INDUSTRIES INC
Manufacturer Address1140 EDISON ST YORK PA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-02-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.