FETAL GUARD FTLGRD 3000 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1998-11-16 for FETAL GUARD FTLGRD 3000 * manufactured by Air Shields - Vickers.

Event Text Entries

[15483777] Pt presented in labor and was assessed in triage using fetal monitor. Fetal monitor strip showed fetal heart tones 150-160. Pt admitted and moved to another room with another fetal monitor. This monitor also showed fetal heart tones at 150-160 with occasional breaks where rate dropped to 70-80's which was assessed as maternal heart rate. Rate always returned to 150's. Physician came in to deliver baby. He ruptured bag of water and applied fetal scalp electrode - no heart rate was attainable by monitor or doptone. An emergency c-section was done with delivery of stillborn male. Autopsy reported baby had been dead in utero for up to 48 hours prior to delivery. Fetal monitors apparently were doubling maternal rate for fetal heart tone. Monitors under warranty and removed from service.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number199239
MDR Report Key199239
Date Received1998-11-16
Date of Report1998-11-16
Date of Event1998-11-03
Date Facility Aware1998-11-03
Date Reported to FDA1998-11-16
Date Reported to Mfgr1998-11-16
Date Added to Maude1998-12-01
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameFETAL GUARD
Generic NameFETAL MONITOR
Product CodeHEL
Date Received1998-11-16
Returned To Mfg1998-11-13
Model NumberFTLGRD 3000
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age*
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key193547
ManufacturerAIR SHIELDS - VICKERS
Manufacturer Address330 JACKSONVILLE ROAD HATBORO PA 190402211 US
Baseline Brand NameFETALGUARD 3000
Baseline Generic NameFETAL MONITOR
Baseline Model NoIP10-E-104
Baseline Catalog NoUNK
Baseline ID*

Device Sequence Number: 2

Brand NameFETALGUARD
Generic NameFETAL MONITOR
Product CodeHEL
Date Received1998-11-16
Returned To Mfg1998-11-13
Model NumberFTLGRD 3000
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age*
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key193548
ManufacturerAIR SHIELDS - VICKERS
Manufacturer Address330 JACKSONVILLE ROAD HATBORO PA 190402211 US
Baseline Brand NameFETALGUARD 3000
Baseline Generic NameFETAL MONITOR
Baseline Model NoIP10-E-104
Baseline Catalog NoUNK
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 1998-11-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.