CONCHATHERM NEPTUNE *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2011-02-14 for CONCHATHERM NEPTUNE * manufactured by Teleflex Inc.

Event Text Entries

[19659026] Upper middle part of the high flow circuit melted and stuck to bed linen. The humidifier did not alarm; temperature reading 28 degrees celsius; circuit felt hot to the touch (felt hotter than 37 degrees c). Condensation noted below the melted part of circuit; very little condensation noted above it. Detached high flow circuit from the nsala cannula and unplugged power plug of neptune humidifier. Administered 02 blow by via bag and mask; patient sat saturation recovered. The melted circuit had been in use for approximately 7 days.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1999606
MDR Report Key1999606
Date Received2011-02-14
Date of Report2011-02-14
Date of Event2011-01-30
Report Date2011-02-14
Date Reported to FDA2011-02-14
Date Added to Maude2011-02-23
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameCONCHATHERM NEPTUNE
Generic NameHEATED HUMIDIFIER
Product CodeBTT
Date Received2011-02-14
Model Number*
Catalog Number*
Lot Number*
ID Number*
OperatorNURSE
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX INC
Manufacturer AddressPO BOX 12600 DURHAM NC 27709 US 27709

Device Sequence Number: 2

Brand NameHEATED WIRE CIRCUIT
Generic NameHEATED WIRE CIRCUIT
Product CodeOFP
Date Received2011-02-14
Model Number*
Catalog Number780-20
Lot Number*
ID Number*
Device Sequence No2
Device Event Key0
ManufacturerTELEFLEX INC HUDSON RCI
Manufacturer AddressPO BOX 12600 DURHAM NC 27709 US 27709

Device Sequence Number: 3

Brand NameCONCHA COLUMN
Generic NameCONCHA COLUMN
Product CodeBTT
Date Received2011-02-14
Model Number*
Catalog Number385-30
Lot Number*
ID Number*
Device Sequence No3
Device Event Key0
ManufacturerTELEFLEX INC HUDSON RCI
Manufacturer AddressPO BOX 12600 DURHAM NC 27709 US 27709


Patients

Patient NumberTreatmentOutcomeDate
10 2011-02-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.