MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2011-02-14 for CONCHATHERM NEPTUNE * manufactured by Teleflex Inc.
[19659026]
Upper middle part of the high flow circuit melted and stuck to bed linen. The humidifier did not alarm; temperature reading 28 degrees celsius; circuit felt hot to the touch (felt hotter than 37 degrees c). Condensation noted below the melted part of circuit; very little condensation noted above it. Detached high flow circuit from the nsala cannula and unplugged power plug of neptune humidifier. Administered 02 blow by via bag and mask; patient sat saturation recovered. The melted circuit had been in use for approximately 7 days.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1999606 |
| MDR Report Key | 1999606 |
| Date Received | 2011-02-14 |
| Date of Report | 2011-02-14 |
| Date of Event | 2011-01-30 |
| Report Date | 2011-02-14 |
| Date Reported to FDA | 2011-02-14 |
| Date Added to Maude | 2011-02-23 |
| Event Key | 0 |
| Report Source Code | User Facility report |
| Manufacturer Link | N |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 0 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Single Use | 0 |
| Previous Use Code | 0 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | CONCHATHERM NEPTUNE |
| Generic Name | HEATED HUMIDIFIER |
| Product Code | BTT |
| Date Received | 2011-02-14 |
| Model Number | * |
| Catalog Number | * |
| Lot Number | * |
| ID Number | * |
| Operator | NURSE |
| Device Availability | N |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | TELEFLEX INC |
| Manufacturer Address | PO BOX 12600 DURHAM NC 27709 US 27709 |
| Brand Name | HEATED WIRE CIRCUIT |
| Generic Name | HEATED WIRE CIRCUIT |
| Product Code | OFP |
| Date Received | 2011-02-14 |
| Model Number | * |
| Catalog Number | 780-20 |
| Lot Number | * |
| ID Number | * |
| Device Sequence No | 2 |
| Device Event Key | 0 |
| Manufacturer | TELEFLEX INC HUDSON RCI |
| Manufacturer Address | PO BOX 12600 DURHAM NC 27709 US 27709 |
| Brand Name | CONCHA COLUMN |
| Generic Name | CONCHA COLUMN |
| Product Code | BTT |
| Date Received | 2011-02-14 |
| Model Number | * |
| Catalog Number | 385-30 |
| Lot Number | * |
| ID Number | * |
| Device Sequence No | 3 |
| Device Event Key | 0 |
| Manufacturer | TELEFLEX INC HUDSON RCI |
| Manufacturer Address | PO BOX 12600 DURHAM NC 27709 US 27709 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2011-02-14 |