MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-02-25 for AU640 CHEMISTRY ANALYZER AU643-02E, AU640 WITH ISE N1256100 manufactured by Beckman Coulter Mishima K.k..
[1930256]
A customer contacted beckman coulter inc. (bci) in regards erroneous low phenytoin result (<2. 5 mg/l) generated on au640 chemistry analyzer. The erroneous result was reported out of the laboratory and questioned by the physician. A fresh pour off sample was retested and resulted in higher result 13. 1 mg/l, which was within the therapeutic range. Patient treatment was not impacted by this event.
Patient Sequence No: 1, Text Type: D, B5
[9101805]
Au640 system using microgenics cedia tdm dilantin (phenytoin). Three level controls before and after the event was in range. Clot was detected at 7:15 am. Service was not dispatched. A root cause for this event it presumed a malfunction in sample aspiration or user error in sample handling.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2050012-2011-00518 |
MDR Report Key | 2001572 |
Report Source | 06 |
Date Received | 2011-02-25 |
Date of Report | 2011-01-25 |
Date of Event | 2011-01-21 |
Date Mfgr Received | 2011-01-25 |
Device Manufacturer Date | 2007-08-01 |
Date Added to Maude | 2011-02-28 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 0 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MS. NORA ZEROUNIAN |
Manufacturer Street | 250 S. KRAEMER BLVD. |
Manufacturer City | BREA CA 92821 |
Manufacturer Country | US |
Manufacturer Postal | 92821 |
Manufacturer Phone | 7149613436 |
Manufacturer G1 | BECKMAN COULTER MISHIMA K.K. |
Manufacturer Street | 454-32 HIGASHINO, NAGAIZUMI-CH SUNTO-GUN |
Manufacturer City | SCHIZUOKA, JP-NOTA 411-0931 |
Manufacturer Country | JA |
Manufacturer Postal Code | 411-0931 |
Single Use | 3 |
Remedial Action | OT |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | AU640 CHEMISTRY ANALYZER |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | DKH |
Date Received | 2011-02-25 |
Model Number | AU643-02E, AU640 WITH ISE |
Catalog Number | N1256100 |
Lot Number | N/A |
ID Number | N/A |
Device Availability | Y |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BECKMAN COULTER MISHIMA K.K. |
Manufacturer Address | 454-32 HIGASHINO, NAGAIZUMI-CH SUNTO-GUN SCHIZUOKA, JP-NOTA 411-0931 JA 411-0931 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2011-02-25 |