MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-02-25 for AU640 CHEMISTRY ANALYZER AU643-02E, AU640 WITH ISE N1256100 manufactured by Beckman Coulter Mishima K.k..
        [1930256]
A customer contacted beckman coulter inc. (bci) in regards erroneous low phenytoin result (<2. 5 mg/l) generated on au640 chemistry analyzer. The erroneous result was reported out of the laboratory and questioned by the physician. A fresh pour off sample was retested and resulted in higher result 13. 1 mg/l, which was within the therapeutic range. Patient treatment was not impacted by this event.
 Patient Sequence No: 1, Text Type: D, B5
        [9101805]
Au640 system using microgenics cedia tdm dilantin (phenytoin). Three level controls before and after the event was in range. Clot was detected at 7:15 am. Service was not dispatched. A root cause for this event it presumed a malfunction in sample aspiration or user error in sample handling.
 Patient Sequence No: 1, Text Type: N, H10
| Report Number | 2050012-2011-00518 | 
| MDR Report Key | 2001572 | 
| Report Source | 06 | 
| Date Received | 2011-02-25 | 
| Date of Report | 2011-01-25 | 
| Date of Event | 2011-01-21 | 
| Date Mfgr Received | 2011-01-25 | 
| Device Manufacturer Date | 2007-08-01 | 
| Date Added to Maude | 2011-02-28 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Health Professional | 0 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 0 | 
| Event Location | 0 | 
| Manufacturer Contact | MS. NORA ZEROUNIAN | 
| Manufacturer Street | 250 S. KRAEMER BLVD. | 
| Manufacturer City | BREA CA 92821 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 92821 | 
| Manufacturer Phone | 7149613436 | 
| Manufacturer G1 | BECKMAN COULTER MISHIMA K.K. | 
| Manufacturer Street | 454-32 HIGASHINO, NAGAIZUMI-CH SUNTO-GUN | 
| Manufacturer City | SCHIZUOKA, JP-NOTA 411-0931 | 
| Manufacturer Country | JA | 
| Manufacturer Postal Code | 411-0931 | 
| Single Use | 3 | 
| Remedial Action | OT | 
| Previous Use Code | 3 | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | AU640 CHEMISTRY ANALYZER | 
| Generic Name | CLINICAL CHEMISTRY ANALYZER | 
| Product Code | DKH | 
| Date Received | 2011-02-25 | 
| Model Number | AU643-02E, AU640 WITH ISE | 
| Catalog Number | N1256100 | 
| Lot Number | N/A | 
| ID Number | N/A | 
| Device Availability | Y | 
| Device Eval'ed by Mfgr | N | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | BECKMAN COULTER MISHIMA K.K. | 
| Manufacturer Address | 454-32 HIGASHINO, NAGAIZUMI-CH SUNTO-GUN SCHIZUOKA, JP-NOTA 411-0931 JA 411-0931 | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2011-02-25 |