[1930256] A customer contacted beckman coulter inc. (bci) in regards erroneous low phenytoin result (<2. 5 mg/l) generated on au640 chemistry analyzer. The erroneous result was reported out of the laboratory and questioned by the physician. A fresh pour off sample was retested and resulted in higher result 13. 1 mg/l, which was within the therapeutic range. Patient treatment was not impacted by this event.
Patient Sequence No: 1, Text Type: D, B5


[9101805] Au640 system using microgenics cedia tdm dilantin (phenytoin). Three level controls before and after the event was in range. Clot was detected at 7:15 am. Service was not dispatched. A root cause for this event it presumed a malfunction in sample aspiration or user error in sample handling.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-00518
MDR Report Key2001572
Report Source06
Date Received2011-02-25
Date of Report2011-01-25
Date of Event2011-01-21
Date Mfgr Received2011-01-25
Device Manufacturer Date2007-08-01
Date Added to Maude2011-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613436
Manufacturer G1BECKMAN COULTER MISHIMA K.K.
Manufacturer Street454-32 HIGASHINO, NAGAIZUMI-CH SUNTO-GUN
Manufacturer CitySCHIZUOKA, JP-NOTA 411-0931
Manufacturer CountryJA
Manufacturer Postal Code411-0931
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAU640 CHEMISTRY ANALYZER
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeDKH
Date Received2011-02-25
Model NumberAU643-02E, AU640 WITH ISE
Catalog NumberN1256100
Lot NumberN/A
ID NumberN/A
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER MISHIMA K.K.
Manufacturer Address454-32 HIGASHINO, NAGAIZUMI-CH SUNTO-GUN SCHIZUOKA, JP-NOTA 411-0931 JA 411-0931