CESAR-OMCP-VISUB (HM3000) 72239

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-01-28 for CESAR-OMCP-VISUB (HM3000) 72239 manufactured by Philips Healthcare.

Event Text Entries

[17895138] The customer reported that the unit has a problem with the image processing which restarts spontaneously.
Patient Sequence No: 1, Text Type: D, B5


[17916687] (b)(4). The investigation found that the customer informed us of a power outage. The field service engineer (fse) reported that the unit has a problem with the digital which restarts spontaneously as a result the disc image was reformatted. In addition, one of the panels in the camera control cabinet fails intermittently, which required replacement of the idm board.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003768277-2011-00084
MDR Report Key2002537
Report Source07
Date Received2011-01-28
Date of Report2011-01-03
Date Mfgr Received2011-01-03
Date Added to Maude2011-03-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANTHONY PERRY
Manufacturer Street595 MINER ROAD
Manufacturer CityCLEVELAND OH 44143
Manufacturer CountryUS
Manufacturer Postal44143
Manufacturer Phone4404832025
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCESAR-OMCP-VISUB (HM3000)
Product CodeIZF
Date Received2011-01-28
Model Number72239
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS HEALTHCARE
Manufacturer AddressVEENPLUIS 4-6 BEST 5680DA NL 5680 DA


Patients

Patient NumberTreatmentOutcomeDate
10 2011-01-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.