UNICEL? DXC 600I SYNCHRON ? CHEMISTRY ANALYZER A27318

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-03-04 for UNICEL? DXC 600I SYNCHRON ? CHEMISTRY ANALYZER A27318 manufactured by Beckman Coulter Inc..

Event Text Entries

[1715060] A customer contacted beckman coulter inc. (bci) in regards to an erroneous high salicylates results on one patient generated by the unicel dxc 600i synchron chemistry analyzer. The erroneous result was reported out of the laboratory. The sample was repeated and lower result was obtained. Amended report was initiated. Patient was admitted to the hospital based on the false high result. Unknown if any treatment was administered.
Patient Sequence No: 1, Text Type: D, B5


[8952672] Qc was run prior to the event and the results were within established ranges. The customer found the sample probe and collar wash contained gel from the serum separator tube. The customer replaced the parts and requested service. A bci field service engineer (fse) performed the followings: cleaned the cartridge chemistry (cc) sample probe and collar wash. Replaced the cc level sense bead and performed alignment. Ran sample and confirmed the correct results.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-00542
MDR Report Key2008151
Report Source06
Date Received2011-03-04
Date of Report2011-02-07
Date of Event2011-02-05
Date Mfgr Received2011-02-07
Device Manufacturer Date2007-08-30
Date Added to Maude2011-03-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNICEL? DXC 600I SYNCHRON ? CHEMISTRY ANALYZER
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeDKJ
Date Received2011-03-04
Model NumberDXC 600I
Catalog NumberA27318
Lot NumberN/A
ID NumberN/A
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2011-03-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.