COULTER LH 750 SLIDE STAINER ANALYZER 6605645

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-03-10 for COULTER LH 750 SLIDE STAINER ANALYZER 6605645 manufactured by Beckman Coulter, Inc..

Event Text Entries

[1915449] A customer contacted beckman coulter inc. (bci) reporting a leak in the lh 750 slide stainer analyzer due to a fluid that was observed over the catch tray. The leak was contained internally within the slide stainer and did not leak outside the instrument. No injuries occurred and medical attention was not sought. There was no report of exposure to mucous membranes or open wounds.
Patient Sequence No: 1, Text Type: D, B5


[9158210] A field service engineer (fse) was dispatched for this event and confirmed that the leak was methanol. The leak was considered minimal and was cleaned up according to the defined laboratory protocol by the fse. The fse replaced the peristaltic pump and verified repair per established procedures. Results met published performance specifications. The root cause for the leak was associated with the peristaltic pump. Per product labeling, bci urges its customers to comply with all national health and safety standards such as the use of barrier protection. This may include, but it is not limited to, protective eyewear, gloves, and suitable laboratory attire when operating or maintaining this or any other automated laboratory analyzer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2011-00126
MDR Report Key2015325
Report Source06
Date Received2011-03-10
Date of Report2010-08-10
Date of Event2010-08-09
Date Mfgr Received2010-08-10
Device Manufacturer Date2004-08-01
Date Added to Maude2012-02-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Remedial ActionRL
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameCOULTER LH 750 SLIDE STAINER ANALYZER
Generic NameAUTOMATED DIFFERENTIAL CELL COUNTER
Product CodeLOQ
Date Received2011-03-10
Model NumberLH 750 SLIDE STAINER
Catalog Number6605645
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address11800 SW 147TH AVENUE MIAMI FL 33196 US 33196


Patients

Patient NumberTreatmentOutcomeDate
10 2011-03-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.