BREMER 3D CROWN SYSTEM 292301001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,07 report with the FDA on 2011-03-16 for BREMER 3D CROWN SYSTEM 292301001 manufactured by Depuy Spine, Inc..

Event Text Entries

[20412936] On (b)(6) 2010, the 3d crown was fixed to the pt. In (b)(6) 2011, it was found through mri that the anterior pins penetrated the inner table of skull. Cerebrospinal fluid leaked from the right forehead bone. Cramp occurred and the level of consciousness was decreased. Also, pneumocephalus occurred together. Then pt's condition was monitored. The pt's previous disease was rheumatism. The doctor requested to know why 8 pound torque is appropriate for the fixation. As an adverse outcome was reported, an mdr is filed to document this event.
Patient Sequence No: 1, Text Type: D, B5


[20577221] Skull pin penetration is a known possible adverse outcome. Pt had rheumatism which indicates that the pt may have compromised bone quality. At this time, the specific pt info is not known. Pin insertion should be done under fluoro to ensure proper penetration. Then the pins are re-tightened 24-48 hours after placement and then as needed. The system comes with torque limiting caps that break away at 8 in/lbs. For pts with compromised bone, the user should use the torque pin driver that comes in the package. The pin driver should also be used when re-tightening to 2-3 in/lbs.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1526439-2011-00039
MDR Report Key2023370
Report Source01,05,07
Date Received2011-03-16
Date Mfgr Received2011-02-16
Device Manufacturer Date2010-09-01
Date Added to Maude2011-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANDREW TOPOULOS
Manufacturer Street325 PARAMOUNT DR.
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088282790
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameBREMER 3D CROWN SYSTEM
Generic NameSKULL BRACE/HAX
Product CodeHAX
Date Received2011-03-16
Model NumberNA
Catalog Number292301001
Lot NumberJ61007AR
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SPINE, INC.
Manufacturer Address325 PARAMOUNT DR. RAYNHAM MA 02767 US 02767


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2011-03-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.