SEDECAL FLEXI-DT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2011-03-14 for SEDECAL FLEXI-DT manufactured by Sedecal S.a. Algete.

Event Text Entries

[1920088] This event involves a powered x-ray table. The operator reportedly rec'd an electric shock while he was connecting the power supply cord to the table. The power cord was already connected to the wall socket. The operator used the cord in a visibly damaged state. The system had been in use for more than 5 years. This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: D, B5


[9092261] The root cause of this event is improper maintenance coupled with user error.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9617251-2011-00001
MDR Report Key2023842
Report Source01,08
Date Received2011-03-14
Date of Report2011-03-08
Date of Event2011-11-24
Date Mfgr Received2011-02-23
Device Manufacturer Date2005-11-01
Date Added to Maude2011-03-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street230 LEXINGTON DR.
Manufacturer CityBUFFALO GROVE IL 60089
Manufacturer CountryUS
Manufacturer Postal60089
Manufacturer Phone8473946960
Manufacturer G1SEDECAL S.A.
Manufacturer StreetPELAYA, 9- POLIGONO INDUSTRIAL
Manufacturer CityRIO DE JANEIRO, MADRIS
Manufacturer CountrySP
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSEDECAL
Generic NameX-RAY TABLE
Product CodeIZZ
Date Received2011-03-14
Model NumberFLEXI-DT
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSEDECAL S.A. ALGETE
Manufacturer AddressMADRID SP


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2011-03-14

© 2026 FDA.report