PRECISION MEDICAL FLOW METER 2MFA1001 2MFA 1001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-03-17 for PRECISION MEDICAL FLOW METER 2MFA1001 2MFA 1001 manufactured by Precision Medical Inc..

Event Text Entries

[1917611] Title: xxxxx. Event desc: nurse noticed something was not right with the flow meter and asked the charge nurse to check the flow meter. The pt was on an oxygen face mask connected to the flow meter involved in the incident. As the charge nurse entered the room, she noticed that it smelled like something was burning. She observed that the flow tube was black. She had respiratory therapy paged to come to the room. She then began to turn down the flow meter, as she was adjusting the flow meter, it exploded and flames began to shoot out of the wall. A code red was initiated via the phone. The pt was immediately evacuated from the room and the door to the room was closed. The nursing supervisor turned off the oxygen to half of the unit. Security extinguished the burning pieces. There was no pt harm, employee injury or damage to the facility.
Patient Sequence No: 1, Text Type: D, B5


[9162746] Notified by hospital that there had been a fire incident with a flow meter in the icu ward on (b)(6) 2011. Notified by hospital on (b)(6) 2011. Visited hospital on (b)(6) 2011. Fire occurred in the icu ward, with a 2mfa flow meter. Flow meter had been operating at 10 liters per minute for several days. Fire occurred in flow meter. Cause is unk, at this time. Flow meter was sent to an outside agency (ecri) for investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2523148-2011-00001
MDR Report Key2026586
Report Source06
Date Received2011-03-17
Date of Report2011-03-01
Date of Event2011-02-24
Date Mfgr Received2011-03-01
Device Manufacturer Date1995-05-01
Date Added to Maude2012-06-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM PARKER
Manufacturer Street300 HELD DR
Manufacturer CityNORTHAMPTON PA 18067
Manufacturer CountryUS
Manufacturer Postal18067
Manufacturer Phone6102626090
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRECISION MEDICAL FLOW METER
Generic NameOXYGEN FLOW METER
Product CodeCAX
Date Received2011-03-17
Model Number2MFA1001
Catalog Number2MFA 1001
Lot Number5/95
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerPRECISION MEDICAL INC.
Manufacturer Address300 HELD DR NORTHAMPTON PA 18067 US 18067


Patients

Patient NumberTreatmentOutcomeDate
10 2011-03-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.