CLOSED KNOT PUSHER N/A 902813

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-03-22 for CLOSED KNOT PUSHER N/A 902813 manufactured by Biomet Sports Medicine, Inc..

Event Text Entries

[18181514] It was reported that patient underwent bankart labral repair procedure on (b)(6) 2011. During the procedure, as the doctor attempted to push a western knot down the labrum, the suture was cut. A competitor's product was used to complete the procedure.
Patient Sequence No: 1, Text Type: D, B5


[18205333] Evaluation of returned device found the radius on the business end of the knot pusher had a sharp edge. This report submitted march 22, 2011.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1825034-2011-00197
MDR Report Key2026871
Report Source07
Date Received2011-03-22
Date of Report2011-02-21
Date of Event2011-02-17
Date Mfgr Received2011-02-21
Device Manufacturer Date2010-05-03
Date Added to Maude2011-04-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. VICTOR RODGERS
Manufacturer Street56 E. BELL DRIVE P.O. BOX 587
Manufacturer CityWARSAW IN 46581
Manufacturer CountryUS
Manufacturer Postal46581
Manufacturer Phone5743713800
Manufacturer G1BIOMET SPORTS MEDICINE, INC.
Manufacturer Street56 E. BELL DRIVE P.O. BOX 587
Manufacturer CityWARSAW IN 46581
Manufacturer CountryUS
Manufacturer Postal Code46581
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLOSED KNOT PUSHER
Generic NamePUSHER, SOCKET
Product CodeHXO
Date Received2011-03-22
Model NumberN/A
Catalog Number902813
Lot Number250210
ID NumberN/A
OperatorPHYSICIAN
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOMET SPORTS MEDICINE, INC.
Manufacturer Address56 E. BELL DRIVE P.O. BOX 587 WARSAW IN 46581 US 46581


Patients

Patient NumberTreatmentOutcomeDate
10 2011-03-22