RENU MULTIPLUS MULTI-PURPOSE SOLUTION

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2011-02-23 for RENU MULTIPLUS MULTI-PURPOSE SOLUTION manufactured by Bausch + Lomb.

Event Text Entries

[17473368] Consumer reported burning and stinging after use of the solution. The solution expired in march 2008. Icp mass spectrometry was performed on another sample of this lot and showed a higher than expected amount of trace iron in the bottle of solution. The source of the iron was identified as one particular lot of one of the raw materials, poloxamine. The product involved is expired and a global recall was completed in 2007 for this lot.
Patient Sequence No: 1, Text Type: D, B5


[17649396] Icp mass spectrometry was performed on another sample of this lot and showed a higher than expected amount of trace iron in the bottle of solution. The source of the iron was identified as one particular lot of one of the raw materials, poloxamine. A global recall was completed in 2007 for this lot.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1313525-2011-00003
MDR Report Key2035249
Report Source04
Date Received2011-02-23
Date of Report2010-12-13
Date Mfgr Received2010-12-13
Device Manufacturer Date2006-03-01
Date Added to Maude2011-04-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTES PROUD
Manufacturer Street1400 NORTH GOODMAN ST.
Manufacturer CityROCHESTER NY 14609
Manufacturer CountryUS
Manufacturer Postal14609
Manufacturer Phone5853388549
Manufacturer G1BAUSCH + LOMB
Manufacturer Street8507 PELHAM ROAD
Manufacturer CityGREENVILLE SC 29615
Manufacturer CountryUS
Manufacturer Postal Code29615
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-0715-2007
Event Type3
Type of Report3

Device Details

Brand NameRENU MULTIPLUS MULTI-PURPOSE SOLUTION
Product CodeLYL
Date Received2011-02-23
Lot NumberGC6037
Device Expiration Date2008-03-31
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBAUSCH + LOMB
Manufacturer AddressROCHESTER NY 14609 US 14609


Patients

Patient NumberTreatmentOutcomeDate
10 2011-02-23