COBAS 6000 E601 MODULE 04745922001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-04-05 for COBAS 6000 E601 MODULE 04745922001 manufactured by Roche Diagnostics.

Event Text Entries

[1883793] The user received a questionable free psa generation 2 result for one patient sample. The initial result was 4. 010 ng/ml and was reported outside the laboratory. The repeat result on the cobas e601 analyzer serial number (b)(4) was 0. 399 ng/ml. The user stated the initial result was not acted upon. No adverse events were alleged regarding the event. The free psa reagent lot number was 15810704. The field service representative could not find a cause. He inspected the valves for a malfunction and could find no problems with the system. To verify the analyzer operation, he ran performance testing with all results within specifications.
Patient Sequence No: 1, Text Type: D, B5


[9187805] A specific root cause could not be identified. No adverse effects were alleged regarding the event.
Patient Sequence No: 1, Text Type: N, H10


[9226865] .
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2011-01868
MDR Report Key2038504
Report Source05,06
Date Received2011-04-05
Date of Report2011-05-05
Date of Event2011-03-16
Date Mfgr Received2011-03-16
Date Added to Maude2011-04-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA RICHEAL CLINE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175213833
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS 6000 E601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeNAF
Date Received2011-04-05
Model NumberNA
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.