MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,04,05,06,07 report with the FDA on 2011-03-24 for POWERPRO STERILE TRANSFER CASE, LARGE PRO3110 manufactured by Conmed Linvatec.
[1885832]
The customer reported the lid was broken at the hinges. There was no report of pt injury and/or adverse outcome. A subsequent review of this complaint found that the sterile barrier may have been compromised.
Patient Sequence No: 1, Text Type: D, B5
[9223518]
The product was not returned to conmed linvatec for eval. Product history shows this device was outside its warranty. A two year review of complaint history for this device showed there have been no serious injuries or deaths related to the reported problem. The product ifu warns: inspect the sterile transfer battery case for any damage, such as: cracks in the case or damaged, broken, or worn o-rings, latch, and/or hinge. If damage is found, do not use. Sterility may be compromised and pt infection may occur.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1017294-2011-00006 |
MDR Report Key | 2038591 |
Report Source | 01,04,05,06,07 |
Date Received | 2011-03-24 |
Date of Report | 2011-03-24 |
Date Mfgr Received | 2009-07-09 |
Device Manufacturer Date | 2008-08-01 |
Date Added to Maude | 2011-04-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | BRENDA JOHNSON |
Manufacturer Street | 11311 CONCEPT BLVD. |
Manufacturer City | LARGO FL 33773 |
Manufacturer Country | US |
Manufacturer Postal | 33773 |
Manufacturer Phone | 7273995515 |
Manufacturer G1 | CONMED LINVATEC |
Manufacturer Street | 11311 CONCEPT BLVD. |
Manufacturer City | LARGO FL 33773 |
Manufacturer Country | US |
Manufacturer Postal Code | 33773 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | POWERPRO STERILE TRANSFER CASE, LARGE |
Generic Name | SAW, POWERED, AND ACCESSORIES |
Product Code | FCO |
Date Received | 2011-03-24 |
Catalog Number | PRO3110 |
Lot Number | BBD77391 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | CONMED LINVATEC |
Manufacturer Address | LARGO FL US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2011-03-24 |