[1885832] The customer reported the lid was broken at the hinges. There was no report of pt injury and/or adverse outcome. A subsequent review of this complaint found that the sterile barrier may have been compromised.
Patient Sequence No: 1, Text Type: D, B5


[9223518] The product was not returned to conmed linvatec for eval. Product history shows this device was outside its warranty. A two year review of complaint history for this device showed there have been no serious injuries or deaths related to the reported problem. The product ifu warns: inspect the sterile transfer battery case for any damage, such as: cracks in the case or damaged, broken, or worn o-rings, latch, and/or hinge. If damage is found, do not use. Sterility may be compromised and pt infection may occur.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017294-2011-00006
MDR Report Key2038591
Report Source01,04,05,06,07
Date Received2011-03-24
Date of Report2011-03-24
Date Mfgr Received2009-07-09
Device Manufacturer Date2008-08-01
Date Added to Maude2011-04-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBRENDA JOHNSON
Manufacturer Street11311 CONCEPT BLVD.
Manufacturer CityLARGO FL 33773
Manufacturer CountryUS
Manufacturer Postal33773
Manufacturer Phone7273995515
Manufacturer G1CONMED LINVATEC
Manufacturer Street11311 CONCEPT BLVD.
Manufacturer CityLARGO FL 33773
Manufacturer CountryUS
Manufacturer Postal Code33773
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePOWERPRO STERILE TRANSFER CASE, LARGE
Generic NameSAW, POWERED, AND ACCESSORIES
Product CodeFCO
Date Received2011-03-24
Catalog NumberPRO3110
Lot NumberBBD77391
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCONMED LINVATEC
Manufacturer AddressLARGO FL US


Patients

Patient NumberTreatmentOutcomeDate
10 2011-03-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.