DC1235 DENTAL CHAIR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2011-02-15 for DC1235 DENTAL CHAIR manufactured by Dental Equipment, Llc (dba Newberg Mfg).

Event Text Entries

[1886412] The link arm assembly which holds the cuspidor to the dental chair broken allowing the cuspidor to fall to the floor.
Patient Sequence No: 1, Text Type: D, B5


[8995142] Newberg mfg recently became aware of failures of the link arm assembly that connects some rear mounted options (cuspidors, assistants instrumentation, and hygiene systems) to the dental chair which may potentially break and fall off the dental chair. Outside lab testing determined that the failure was due to hydrogen embrittlement of the link arm bolt. This failure is caused by not properly baking the bolts after they are zinc plated in a timely manner. All bolt failures were from the same lot number. A decision was made by dental equipment llc to conduct a voluntary recall of the affected link arm assemblies.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004517290-2011-00004
MDR Report Key2040306
Report Source08
Date Received2011-02-15
Date of Report2011-02-24
Date of Event2011-02-24
Date Mfgr Received2011-02-24
Device Manufacturer Date2010-11-01
Date Added to Maude2012-02-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY, MANAGER
Manufacturer Street11727 FRUEHAUF DR.
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1DENTAL EQUIPMENT, LLC, DBA NEWBERG MFG
Manufacturer Street705 S. SPRINGBROOK RD. BLDG C
Manufacturer CityNEWBERG OR 97132
Manufacturer CountryUS
Manufacturer Postal Code97132
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDC1235 DENTAL CHAIR
Generic NameDENTAL CHAIR
Product CodeKLC
Date Received2011-02-15
Model NumberDC1235
Catalog NumberDC1235
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDENTAL EQUIPMENT, LLC (DBA NEWBERG MFG)
Manufacturer Address705 SOUTH SPRINGBROOK RD. BLDG C NEWBERG OR 97132 US 97132


Patients

Patient NumberTreatmentOutcomeDate
10 2011-02-15