IMMAGE? MICROALBUMIN (MA) REAGENT N/A 447690

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06 report with the FDA on 2011-04-07 for IMMAGE? MICROALBUMIN (MA) REAGENT N/A 447690 manufactured by Beckman Coulter Inc..

Event Text Entries

[1885970] A customer contacted beckman coulter inc. (bci) stating that the results of controls and patient specimens are unstable for the microalbumin reagent. The customer also stated that the microalbumin reagent generated different results when loaded on different cartridges. No sample information was provided except that sample results were different with different lots or cartridges of reagent. No affect to patients were reported, as the results were not reported out of the laboratory.
Patient Sequence No: 1, Text Type: D, B5


[9178549] No sample information was provided. The sample type for microalbumin is usually random urine. Customer noted that qc has been unacceptable. Customer did not note any system errors or issues. It is unknown whether a new lot or cartridge of reagent resolved the issue. Service was not needed for this event, as this is a reagent issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-01041
MDR Report Key2044583
Report Source01,06
Date Received2011-04-07
Date of Report2011-03-09
Date of Event2011-03-09
Date Mfgr Received2011-03-09
Device Manufacturer Date2010-10-01
Date Added to Maude2011-05-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? MICROALBUMIN (MA) REAGENT
Generic NameMICROALBUMIN (MA) REAGENT
Product CodeJIQ
Date Received2011-04-07
Model NumberN/A
Catalog Number447690
Lot NumberM010304
Device Expiration Date2012-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BOULEVARD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-07

© 2026 FDA.report