WELL LEG HOLDER 72200632

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-04-11 for WELL LEG HOLDER 72200632 manufactured by Smith & Nephew Endoscopy Andover Manufacturing.

Event Text Entries

[1863483] The patient was under anesthesia, the foot piece on the leg holder cracked and broke off during the procedure. Case had to be cancelled.
Patient Sequence No: 1, Text Type: D, B5


[9177071] Returned device was sent to vendor for evaluation, pending investigation result. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[18736656] The product was sent to (b)(4) medical for evaluation and confirm the customer complaint for the foot piece breaking off. The shaft pivot pin has had the head pulled thru the foot plate. A corrective action was opened and previously completed to address this issue. This product was manufactured in (b)(6) 2008 and predates the design change. The eco was approved on (b)(6) 2009 and the fmea was implemented on (b)(6) 2009. No further investigation is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003604053-2011-00006
MDR Report Key2048367
Report Source07
Date Received2011-04-11
Date of Report2011-03-15
Date of Event2011-03-15
Date Facility Aware2011-03-15
Date Mfgr Received2011-03-15
Device Manufacturer Date2008-09-01
Date Added to Maude2011-10-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJASON BILOBRAM
Manufacturer Street150 MINUTEMAN ROAD
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal01810
Manufacturer Phone9787491630
Manufacturer G1ANDOVER MANUFACTURING SITE
Manufacturer Street150 MINUTEMAN ROAD
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal Code01810
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWELL LEG HOLDER
Generic NameWELL LEG HOLDER
Product CodeJEB
Date Received2011-04-11
Returned To Mfg2011-03-25
Model Number72200632
Catalog Number72200632
Lot Number329827
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age30 MO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW ENDOSCOPY ANDOVER MANUFACTURING
Manufacturer Address150 MINUTEMAN ROAD ANDOVER MA 01810 US 01810


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.