COULTER LH 750 ANALYZER 6605632

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-04-14 for COULTER LH 750 ANALYZER 6605632 manufactured by Beckman Coulter Inc..

Event Text Entries

[15408222] A customer contacted beckman coulter inc. (bci) to report that the stoppers from hemogard tubes are pulled off during automatic patient sampling on the coulter lh750 hematology analyzer, resulting in blood spills. User was wearing personal protective equipment (ppe). No injuries occurred and no one sought medical attention. There was no report of exposure to mucous membranes or open wounds. Patient was re-drawn for analysis as the sample was lost when the stopper came out during piercing. No reports of death, injury, or change to patient treatment have been reported in connection with this event.
Patient Sequence No: 1, Text Type: D, B5


[15646575] Account runs approximately 350 patient samples per day, and approximately 10,500 patient samples a month. Customer uses bd 13 x 75 hemogard tubes, which are approved on the tube list. Account occasionally removes the stopper to obtain a sample aliquot for sedimentation sed rates and / or to check for clots before processing on the lh750 instrument. The stopper was not removed from this particular specimen before processing. Bci field service engineer (fse) was dispatched for this event. Needle, rocker bed and surrounding areas were cleaned with disinfectant. Fse verified the rocker bed operation and ran qc with satisfactory results. Per labeling, beckman coulter, inc. Urges its customers to comply with all national health and safety standards such as the use of barrier protection. This may include, but it is not limited to, protective eyewear, gloves, and suitable laboratory attire when operating or maintaining this or any other automated laboratory analyzer. The root cause for the stoppers being pulled off, resulting in a blood spill, is unknown.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2011-00253
MDR Report Key2058258
Report Source05,06
Date Received2011-04-14
Date of Report2011-03-15
Date of Event2011-03-15
Date Mfgr Received2011-03-15
Device Manufacturer Date2006-03-01
Date Added to Maude2012-02-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameCOULTER LH 750 ANALYZER
Generic NameAUTOMATED DIFFERENTIAL CELL COUNTER
Product CodeLOQ
Date Received2011-04-14
Model NumberLH 750
Catalog Number6605632
Lot NumberN/A
ID NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address11800 SW 147TH AVENUE MIAMI FL 33196 US 33196


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-14