MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-04-19 for JETER PHARYNGEAL DISSECTOR 02-0027 manufactured by Biomet Microfixation.
[1941735]
It was reported during a procedure for a zygomatic complex fracture and repair of right orbital floor, 2mm tip of device broke off into soft tissue underneath the bone. At this time, unable to retrieve.
Patient Sequence No: 1, Text Type: D, B5
[9116251]
Visual review of instrument shows dings and gashes on both working ends. Current information is insufficient to permit a valid conclusion about the cause of this event. If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent. Hospital submitted mdr also, see triage unit sequence (b)(4).
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1032347-2011-00054 |
MDR Report Key | 2061863 |
Report Source | 05,06 |
Date Received | 2011-04-19 |
Date of Report | 2011-03-31 |
Date of Event | 2011-03-21 |
Date Mfgr Received | 2011-03-31 |
Date Added to Maude | 2011-04-19 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | RISK MANAGER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MRS. CHRISTY CAIN |
Manufacturer Street | 1520 TRADEPORT DRIVE |
Manufacturer City | JACKSONIVLLE FL 32218 |
Manufacturer Country | US |
Manufacturer Postal | 32218 |
Manufacturer Phone | 9047419468 |
Manufacturer G1 | BIOMET MICROFIXATION |
Manufacturer Street | 1520 TRADEPORT DRIVE |
Manufacturer City | JACKSONVILLE FL 32218 |
Manufacturer Country | US |
Manufacturer Postal Code | 32218 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | JETER PHARYNGEAL DISSECTOR |
Generic Name | DISSECTOR |
Product Code | GDI |
Date Received | 2011-04-19 |
Returned To Mfg | 2011-04-08 |
Catalog Number | 02-0027 |
Lot Number | UNKNOWN |
Operator | PHYSICIAN |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMET MICROFIXATION |
Manufacturer Address | 1520 TRADEPORT DRIVE JACKSONVILLE FL 32218 US 32218 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2011-04-19 |