MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-04-07 for CLINITEK STATUS ANALYZER CT STATUS manufactured by Siemens Healthcare Diagnostics, Manufacturing Ltd..
[1937255]
Pt was admitted due to pelvic pain. Customer reported a false positive urine hcg which was confirmed with serum beta hcg. Pt treatment was delayed until false positive confirmed. No unnecessary medical procedure was performed. No treatment was withheld as a result of this incident. There was no impact to pt health.
Patient Sequence No: 1, Text Type: D, B5
[9175709]
Customer could not confirm whether the control, test or reference lines on the test cassette were visibly developed. Customer did not know when calibration bar had last been cleaned. Controls were last run between (b)(6). Customer was advised to clean calibration bar and run controls on the instrument. Both levels of control passed after cleaning.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1217157-2011-00007 |
MDR Report Key | 2062756 |
Report Source | 07 |
Date Received | 2011-04-07 |
Date of Report | 2011-03-15 |
Date of Event | 2011-03-12 |
Date Mfgr Received | 2011-03-15 |
Device Manufacturer Date | 2005-12-01 |
Date Added to Maude | 2012-06-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MARK SMITH |
Manufacturer Street | 2 EDGEWATER DR. |
Manufacturer City | NORWOOD MA 02062 |
Manufacturer Country | US |
Manufacturer Postal | 02062 |
Manufacturer Phone | 7812693158 |
Manufacturer G1 | SIEMENS HEALTHCARE DIAGNOSTICS, MANUFACTURING LTD. |
Manufacturer Street | NORTHERN RD. |
Manufacturer City | SUDBURY, SUFFOLK CO102XQ |
Manufacturer Country | UK |
Manufacturer Postal Code | CO10 2XQ |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CLINITEK STATUS ANALYZER |
Generic Name | CLINITEK STATUS ANALYZER |
Product Code | LCX |
Date Received | 2011-04-07 |
Model Number | CT STATUS |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS, MANUFACTURING LTD. |
Manufacturer Address | NORTHERN RD. SUDBURY, SUFFOLK CO106DX UK CO10 6DX |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2011-04-07 |