LEKSELL GAMMA KNIFE MODEL 4C NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,07 report with the FDA on 2011-04-05 for LEKSELL GAMMA KNIFE MODEL 4C NA manufactured by Elekta Instrument Ab.

Event Text Entries

[1940353] On (b)(6) 2011, the physicist was trying to get the helmet hoist to lock on the helmet. The hoist down position sensor was failing, so the helmet was not locking. He reported that the hoist was not going to the fully up position also. While trying to drive the hoist down, without a helmet, the actuator snapped, failed, and subsequently caused the helmet hoist to break at the pivot point. The cartridge assy white block snapped around the pivots and fell to the ground.
Patient Sequence No: 1, Text Type: D, B5


[9063711] Investigation is on-going so f/u reporting to this mdr will be provided to fda upon completion.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9612186-2011-00002
MDR Report Key2063796
Report Source00,07
Date Received2011-04-05
Date of Report2011-04-04
Date of Event2011-03-21
Date Mfgr Received2009-06-02
Date Added to Maude2011-07-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetKUNGSTENSGATAN 18
Manufacturer CitySTOCKHOLM
Manufacturer CountrySW
Manufacturer Phone293654250
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEKSELL GAMMA KNIFE
Generic NameSYSTEM, RADIATION THERAPY, RADIONUCLIDE
Product CodeIWB
Date Received2011-04-05
Model NumberMODEL 4C
Catalog NumberNA
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerELEKTA INSTRUMENT AB
Manufacturer AddressKUNGSTENSGATAN 18 STOCKHOLM SW


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-05

© 2026 FDA.report